The Radiation Protection Inspectorate has determined during an extraordinary inspection at the Clinical Center of Montenegro that the reports on dosimetric testing, work environment control and quality control for 32 sources of ionizing radiation are not valid. During the inspection in early September, the institution did not submit to the inspectorate all permits for radiation activities, nor reports on individual exposure of employees to radiation.
Dosimetric tests are measurements that check how much ionizing radiation a device produces and whether people and the area around it are exposed to a higher dose than allowed.
The Inspectorate has given KCCG a deadline of 60 days to submit valid reports on dosimetric testing, work environment control and measurements for the purpose of implementing the quality assurance and control program for all radiation sources used. Within the same deadline, according to the inspection findings, which were reviewed by "Vijesti", the employer is required to arrange and document the health status of all professionally exposed persons for whom it has not been validly determined, including persons with expired findings, incomplete medical examinations, as well as persons with established temporary or permanent disability, and to submit evidence thereof to the Inspectorate.
By October 10, KCCG must submit decisions on the fulfillment of the conditions for performing radiation activities for all sources of ionizing radiation, as well as proof of the radiation dose received by employees who work professionally in such an environment.
The KCCG responded to "Vijesti" that it is more than obvious that the Occupational Health and Safety Service, or rather its employees, led by the then managers and chiefs, have not done their job adequately and properly for several years.
The institution claims that the annual inspection of all sources of ionizing radiation was supposed to be carried out in July 2026, and that at the end of that month they sent a request to the Center for Ecotoxicological Testing, as the authorized institution.
Inconsistencies
The KCCG explains that measurements are carried out regularly, once a year, as prescribed by law. The response states that the inspector found discrepancies between the data contained in the reports on personal dosimeter readings and the actual condition of employees who performed work in the ionizing radiation zone during the period of personal dosimetric control.
"Records on the borrowing, use, surrender and reading of personal dosimeters were not kept properly and up-to-date, i.e. a reliable connection between the specific dosimeter and the person who used it in the analyzed period was not ensured, and CETI performed analyses based on the dosimeters that were delivered to them by the Occupational Safety and Health Service, and for this reason the results in the aforementioned reports are not valid for the inspector. Due to the aforementioned discrepancies, the inspector could not determine whether the dose value read corresponded to the actual individual exposure of the specific occupationally exposed person," the response states.
This method of record keeping, according to the KCCG, has made it impossible to reliably monitor individual professional exposure to ionizing radiation and represents a deficiency in the system of personal dosimetric control.
"Based on the fact that KCCG employs a large number of professionally exposed persons, and a large number of organizational units where professionally exposed persons work, as well as the fact that for years the Occupational Safety and Health Service, which is responsible for performing these tasks according to the systematization, has not kept records adequately, and in accordance with the Law on Protection from Ionizing Radiation and Radiation Safety, the inspector ordered the measure to submit a new report on the examination of the degree of individual exposure of professionally exposed persons by measuring the personal dose equivalent Hp(10) by thermoluminescent dosimetry with current and accurate data," the institution pointed out.
KCCG: Scheduled by the competent service
They claim that the Occupational Safety and Health Service did not regularly submit personal dosimeters to CETI, and also did not keep accurate records of personal dosimeters and persons who used them for a long time, and for this reason, the inspector ordered a new report to be submitted, which will be prepared after the Occupational Safety and Health Service submits every personal dosimeter used in KCCG for reading, with accurate data on who uses the dosimeters.
They said that it was precisely because of this, and also because of the analysis of the work of the Occupational Safety and Health Service, that they made personnel changes in the first act.
"The inspector's report also referred to the medical examinations of persons working in the radiation zone. He determined that the records for the supervisory period 2024, 2025 and 2026 were not in order, again due to the fact that the Occupational Safety and Health Service did not adequately perform that part of the work," the KCCG pointed out.
Asked whether the institution can guarantee to patients and employees that the radiation dose does not exceed the permitted limits, the KCCG said that last year's measurements showed that all devices used at the institution meet positive legal requirements for use, and therefore there is no danger of negative effects of ionizing radiation on employees and patients.
"This year's measurements of all ionizing radiation sources are currently in procedure. Given the large number of ionizing radiation sources used in KCCG, CETI is not physically able to perform measurements in a short period of time, but will complete them as soon as possible," they said.
Disorganized records
They claim that the level of individual exposure of professionally exposed persons is tested regularly by measuring the personal dose equivalent Hp(10) by dosimetry. The measurements, they claim, have shown that none of the analyzed dosimeters registered a value of Hp(10) that could lead to exceeding the maximum permitted value of 20 mSv for the effective dose on an annual basis, regulated by Article 3 of the Regulation on Exposure Limits to Ionizing Radiation (“Official Gazette of the Republic of Yugoslavia” No. 32/98).
"The inspector's statement referred to the fact that he had determined discrepancies between the data contained in the reports on the readings of personal dosimeters and the actual situation of employees who, during the period of personal dosimetric control, performed work in the ionizing radiation zone. Records on the borrowing, use, surrender and reading of personal dosimeters were not kept regularly and up-to-date, i.e. a reliable connection between a specific dosimeter and the person who used it during the analyzed period was not ensured, and CETI performed analyses based on the dosimeters that were delivered to them by the Occupational Safety and Health Service, and for this reason the results in the aforementioned reports cannot be valid," the institution explained.
Due to these discrepancies, they claim, the inspector could not determine whether the dose value read corresponded to the actual individual exposure of the specific occupationally exposed person.
"This method of record keeping has made it impossible to reliably monitor individual occupational exposure to ionizing radiation and represents a deficiency in the personal dosimetric control system," the KCCG said.